5100HD TDLAS

SKU: Model 5100 HD Categories: ,

5100HD TDLAS hosts a dual laser platform. Two separate lasers are installed in the hosts a dual laser platform. Two separate lasers are installed in the instrument, sharing a common optical path through the sample. It enables the measurement of two analytes simultaneously. This results in significant cost savings to our customers. Where the required concentrations of the analytes differ considerably such that different path lengths are required, two sample cells are used.  It is also capable of dual stream analysis. Model 5100HD can measure any two combinations of water vapor, methane, hydrogen sulfide, carbon dioxide, and oxygen in process or flue gas.

Designed to handle sample gases with high dew points, the standard version has its sample cell heated to a maximum of 100°C. A version is also available for 150°C.

A key feature of the Model 5100HD is the use of sealed reference cells, which contain known amounts of analyte gases measured, for referencing the emission wavelengths of the lasers.  The use of reference cells enables both the ability to line-lock the lasers and to continuously verify the instrument calibration.

ETA Process Instrumentation

sales@etapii.com
119 Foster Street, Bldg. #6,
Peabody, MA 01960

Tel: (978) 532-1330
Fax: (978) 532-7325

Visit Our Sister Companies

Scroll to Top